GB CL TSH
High-precision quantitative CLIA assay for TSH in serum/plasma. 20-minute reaction with a 0.015 µIU/mL detection limit. Essential for diagnosing primary, pituitary, or hypothalamic thyroid disorders.
PRODUCT DESCRIPTION
This reagent is an in-vitro diagnostic quantitative assay, using chemiluminescent immunoassay (CLIA) sandwich format, for measuring human TSH in serum or plasma. It requires 80 µL sample volume and delivers results in 20 minutes. With a detection limit of 0.015 µIU/mL, it detects very low TSH concentrations. Stored at 2-8°C, avoid light exposure and freezing. Ideal for diagnosing primary, pituitary, or hypothalamic causes of thyroid dysfunction and monitoring recovery or therapy.
PERFORMANCE
| Limit of Quantitation (LoQ): 0.015 μIU/mL | ||
|---|---|---|
| Precision | Variability | Coefficient of variation, CV (%) |
| Repeatability | within run | <10% |
| Reproducibility | between-run/between-day | <10% |
| between-sites/between-operators | <10% | |
| between-lots | <10% | |
SPECIFICATION
| Principle | Sandwich-type chemiluminescent immunoassay |
| Assay format | Quantification |
| Rreaction time | 20 minutes |
| Sample type | Serum/Plasma |
| Sample volume | 80 μL |
| Detection unit | Relative light unit (RLU) |
| Storage | 2-8°C. Avoid exposure to direct sunlight. Do not freeze. |
| Certification | TFDA |
ORDERING INFORMATION
| Catalogue No. | Product |
|---|---|
| 4ATSH02 | GB CL TSH |
| 4ATSH1B | GB CL TSH-Control |
| 4ATSH2B | GB CL TSH-Calibrator |