Dynex DSX® Automated ELISA with Confidence

The Dynex DSX® is a modular ELISA system with 4 incubators, walk-away automation, reagent protection, sample-plate tracking, and flexible throughput (strip or full-plate) for GB ELISA kits in medium volume labs.

PRODUCT DESCRIPTION

The Dynex DSX® ELISA System is a fully automated, modular ELISA platform designed for medium-throughput laboratories that demand reliable walk-away operation. With four independent plate incubators, the system supports both 1 assay per strip and full 12-assay plates, helping manage workload and reduce labor demands.

Built-in features include reagent light protection (dark cover), barcode sample and plate tracking, “wash buffer low” and “waste full” alarms, fluid-level and tip detection, and error recovery capabilities. These validation and QC features safeguard assay integrity and reproducibility.

Compatible with all GB brand ELISA kits, it offers flexible sample loading, consistent plate temperature control, multiple wash buffer management, and clotted sample protection. Ideal for ELISA users in clinical labs, blood screening, and other institutions where throughput, accuracy, and automation are priorities.

FEATURE

Efficient, Dependable, and Economical

SPECIFICATION

Physical Specification
Dimensions (W x D x H)106 x 91 x 80 cm
Weight136 kg
Power Supply
Input PowerAC 100-240 V, 50/60 Hz, <800 VA
System Specifications
Number of Plates4
Sample Capacity96
Continuous Sample LoadingYes
Sample Tube Size10 mm – 16 mm diameter 40mm – 100 mm height
Number of ReagentsUp to 24 (25 mL tubes)
Number of Standard/Control Bottles33 (1.5 mL vials)
Assays per Plate1 assay per strip or up to 12 assays per plate
Dilution Wells CapacityUp to 192 deep-well strips (24 x 8)
Sample Tip Capacity432 (4 x 108)
Reagent Tip Capacity41 tips
Reader Specifications
Photometric Range0.000 – 3.000 OD
Spectral Range405 nm – 690 nm
Filter Slots6 wavelength
Washer Specifications
Manifold Configuration8-way
Programmable Volumes50 μL – 999 μL
Wash Buffer Capacity4 wash bottles at 2.0 L, with level-sensing
Incubator Specifications
Number of Incubators4
Temperature RangeRT + 7° C to 50° C (elevated incubator) RT + 5° C (ambient incubator)
Temperature Accuracy± 1° C
Shaking> 15 Hz periodic or continuous
Reagent Pipetting Specifications
Reagent Tip Size1,300 μL
Reagent Pipetting Volume25 μL – 1,000 μL
Sample Pipetting Specifications
Sample Tip Size300 μL
Sample Pipetting Volume10 μL – 250 μL single-shot 25 μL – 100  μL multi-shot

ORDERING INFORMATION

Catalogue No.Product
S00338Dynex DSX®